VOLTAGE MEASUREMENTS FOR DISTORTED WAVEFORMS

H. Ndilimabaka, P. Espel, A. Poletaeff
  • Abstract:
    This paper describes a standard system at LNE to measure RMS value of distorted voltage waveforms by means of a sampling voltmeter. Thermal converters are used to trace this quantity to the SI units. Deviations within 15 μV/V are obtained whatever the total harmonic distortion of the waveforms.
  • Keywords:
  • DOI:
    _unreg_tc19-2010.005

Event details:

  • IMEKO TC:
    TC19
  • Event name:
    Instrumentation for the Information and Communication Technology Era
  • Title:
    3rd IMEKO TC19 Symposium (together with 17th IMEKO TC4 Symposium & 15th International Workshop on ADC Modelling and Testing)
  • Place:
    Kosice, SLOVAKIA
  • Time:
    08 September 2010 - 10 September 2010