Search Proceedings
Who we are
Use of a 100 MΩ step Hamon resistor for calibration of pico-ammeters in dc current in the range 100 pA ÷ 100 nA
Pier Paolo Capra, Flavio Galliana, Milena Astrua, E. Gasparotto
-
Abstract:This paper describes a measurement circuit for the calibration of pico-ammeters in dc current. The system is based on a 100 MΩ Hamon network developed at INRIM to improve the traceability level of the maintained 1 GΩ standard. Besides the Hamon network, the measurement system, presented in this paper, consists of a precision dc voltage source and an electronic circuit used as voltage guard driver.
-
Keywords:
-
Download:
-
DOI:_unreg_tc4-2007.012
Event details:
-
IMEKO TC:TC4
-
Event name:TC4 Symposium 2007
-
Title:XVth IMEKO TC4 International Symposium on Novelties in Electrical Measurements and Instrumentation (together with 12th IMEKO TC4 Workshop on ADC Modelling and Testing)
-
Place:Iasi, ROMANIA
-
Time:19 September 2007 - 21 September 2007