Skip to main content

UNCERTAINTY ANALYSIS OF CELL COUNTING BY METABOLIC ASSAYS

C.Divieto, G.Sassi, M.P.Sassi
  • Abstract:
    Cell counting is an important procedure in cell biology and its uncertainty evaluation is fundamental to improve cell-based measurement comparability which is currently variable. In this study, for the first time, the uncertainty of a metabolic assay for cell counting based on fluorescence intensity measurement has been evaluated. Results, in 2D cell culture, revealed that three are the main components: repeatability and reproducibility of the measurement system and a contribution due to the sample loading step procedure. This study represents the basis of a 3D cell culture in which metabolic methodologies are the most suitable method to monitor cell growth.
  • Keywords:
    2D and 3D cell culture, scaffold, cell proliferation rate, CellTiter Blue, resazurin, uncertainty analysis
  • DOI:
    _unreg_wc-2012.SS2-O11

Event details:

  • IMEKO TC:
  • Event name:
    XX IMEKO World Congress
  • Title:

    Metrology for Green Growth

  • Place:
    Busan, REPUBLIC of KOREA
  • Time:
    09 September 2012 - 12 September 2012