Search Proceedings
Who we are
Traceable Low Frequency AC Voltage Measurements
Christiaan S. Veldman
-
Abstract:The International Organization for Standardization (ISO) has developed a set of standards for the calibration of vibration and shock transducers. Method 3 of ISO 16063 part 11, deals specifically with the primary calibration of accelerometers over the frequency range 1 Hz to 10 kHz. Many NMIs have implemented the method over a wider frequency range. One of the difficulties faced by said NMI, is the provision of traceable low frequency AC voltage measurements.
This paper considers low frequency AC voltage traceability options and investigates in more detail a measurement option defined as DC sampled AC measurements. As part of the investigation; 1) a calibration procedure is developed and implemented, 2) the measurement accuracy of the system is demonstrated. -
Keywords:ISO 16063, vibration transducer, calibration standard, AC voltage, calibration, traceability
-
Download:
-
DOI:_unreg_tc22-2007.110
Event details:
-
IMEKO TC:TC22
-
Event name:Cultivating metrological knowledge
-
Title:1st Conference on Vibration Measurement (together with 20th TC3 Conference on Measurement of Force, Mass and Torque & 3rd TC16 Conference on Pressure Measurement)
-
Place:Merida, MEXICO
-
Time:27 November 2007 - 01 December 2007