Search Proceedings
Who we are
THE SETUP AND PERFORMANCE EVALUATION OF ACCELEROMETER COMPARISON CALIBRATION SYSTEM IN NML TAIWAN
Tsung Hsien Tu, Kuang-Yih Tsuei, Jiun-Kai Chen, Yu-Chung Huang, Pei-Yao Yu, Kuang-I Chang
-
Abstract:Accelerometer comparison calibration technique involves the back-to-back coupling the accelerometer under test directly to a traceable reference standard accelerometer and both accelerometers will be experienced the same motion from shaker. Thus the sensitivity of standard accelerometer can be transferred to the test accelerometer via a calibration constant from the output signals. The results of the relative expand uncertainty for the voltage sensitivities for accelerometer are less than 1.4 %, 1.9 % and 2.9 % within frequency range from 50 Hz to 1.5 kHz, from 1.5 kHz to 5 kHz, and from 5 kHz to 7 kHz, respectively. This paper will present the construction, the performance evaluation of the calibration system in NML, CMS/ITRI.
-
Keywords:accelerometer calibration, ISO 16063-21
-
Download:
-
DOI:_unreg_wc-2015.442
Event details:
-
IMEKO TC:
-
Event name:XXI IMEKO World Congress
-
Title:
Measurement in Research and Industry
-
Place:Prague, CZECH REPUBLIC
-
Time:30 August 2015 - 04 September 2015