Skip to main content

THE COMMISSIONING OF THE FIRST UK NATIONAL STANDARD STATIC TORQUE CALIBRATION MACHINE

Andy Robinson
  • Abstract:
    At IMEKO - XV World Congress, the design of the first UK national torque calibration machine was described. Following the construction of the machine, this paper describes its commissioning, leading to the establishment of the first UK national static torque calibration standard. The novel vertical transducer orientation is made possible through several innovative sub-assemblies. This paper evaluates the performances of these critical sub-assemblies, and compares them against design specifications.
  • DOI:
    _unreg_tc3-2005.052

Event details:

  • IMEKO TC:
    TC3
  • Event name:
    Force, Mass and Torque Measurements
  • Title:
    Theory and Applications in Laboratories and Industry
  • Place:
    Cairo, EGYPT
  • Time:
    19 February 2005 - 23 February 2005