Search Proceedings
Who we are
REALISATION OF A PRIMARY AIR VELOCITY STANDRAD USING LASER DOPPLER ANEMOMETER AND PRECISION WIND TUNNEL
Jian Wu
-
Abstract:A primary air velocity meter calibration standard comprising of a precisely controlled wind tunnel system and a Laser Doppler Anemometry (LDA) system was developed for the calibration of air velocity meters from 0.2 m/s to 60 m/s with measurement uncertainties of less than 0.18% to 0.63%. The LDA system is used to verify the wind speed in the test section of the wind tunnel, and provide traceability to the SI for air speed measurements.
-
Keywords:air velocity measurement, wind profiles, measurement uncertainty
-
Download:
-
DOI:_unreg_wc-2009.413
Event details:
-
IMEKO TC:
-
Event name:XIX IMEKO World Congress
-
Title:
Fundamental and Applied Metrology
-
Place:Lisbon, PORTUGAL
-
Time:6 September 2009 - 11 September 2009