Search Proceedings
Who we are
PTB AND INRIM HIGH FORCE INTERCOMPARISON UP TO 9 MN
C. Ferrero, C. Marinari, R. Kumme, W. Herte
-
Abstract:The present paper describes the long term stability of the main metrological characteristics of the two reference transducers of 3 MN and 9 MN together with the relevant uncertainties evaluated from the measurements carried out by INRIM and PTB by bilateral comparisons. In June 2004, the new INRIM – 3 MN build-up system (BU), the 9 MN and 5 MN transducers were also intercompared, by using the PTB 2 MN DWM and the 16.5 MN Hydraulic Multiplication Machines (HMM). The main results are reported. Furthermore long term investigations carried out over 20 years and the work obtained within a EUROMET project are described.
-
Keywords:force standard machine, long term stability, intercomparison
-
Download:
-
DOI:_unreg_wc-2006.TC3-002
Event details:
-
IMEKO TC:
-
Event name:XVIII IMEKO World Congress
-
Title:
Metrology for a Sustainable Development
-
Place:Rio de Janeiro, BRAZIL
-
Time:17 September 2006 - 22 September 2006