Search Proceedings
Who we are
PROFICIENCY TEST AMONG CENTRIFUGE CALIBRATION LABORATORIES
Akihiro Ota, Hideaki Nozato, Wataru Kokuyama, Sakae Chiba, Masao Shimizu, Hiroyuki Kudo, Kenji Maki, Ryuuichi Tada, Chihiro Mochizuki, Shouji Iso
-
Abstract:A proficiency test was carried out to check proficiency of centrifuge calibration among four participants in Japan; KYOWA, TTDC, JARI and NICS, as a pilot laboratory of AIST. AIST provides reference value to the test by using primary shock calibration system. As a result, the proficiency for all participants was evaluated by using z-scores. The z value for respective participants is estimated to be below 2. This means all participants have enough proficiency for centrifuge calibration.
-
Keywords:safety, car crash test, shock calibration, centrifuge calibration, piezoresistive accelerometer
-
Download:
-
DOI:_unreg_wc-2015.436
Event details:
-
IMEKO TC:
-
Event name:XXI IMEKO World Congress
-
Title:
Measurement in Research and Industry
-
Place:Prague, CZECH REPUBLIC
-
Time:30 August 2015 - 04 September 2015