Search Proceedings
Who we are
PRIMARY VACUUM STANDARD CMI FOR UHV PRESSURE RANGE 1 × 10-6 TO 1 × 10-10 Pa
T. Gronych, L. Peksa, M. Jeráb, Z. Krajícek, D. Pražák, F. Stanek, M. Vicar, J. Tesar
-
Abstract:A primary vacuum standard for UHV pressure range is under development in Czech Metrology Institute (CMI). In order to cover the UHV range down to the limit 1 × 10-10 Pa a method of calibration in non-equilibrium gas (so called dynamic extension method) was employed. The apparatus is divided into three parts: the dynamic extension stage, the orifice flow standard and the flowmeter. The dynamic extension stage is designed as an XHV apparatus manufactured from a low outgassing rate material – beryllium copper alloy. A special 4 K bakeable refrigerator cryopump is used as the main pump.
-
Keywords:primary vacuum standard, vacuum gauge calibration in UHV range
-
Download:
-
DOI:_unreg_wc-2012.TC16-P4
Event details:
-
IMEKO TC:
-
Event name:XX IMEKO World Congress
-
Title:
Metrology for Green Growth
-
Place:Busan, REPUBLIC of KOREA
-
Time:09 September 2012 - 12 September 2012