Skip to main content

PREPARATION PROCEDURE FOR LOADING FLUX QUANTUM WEIGHTS ON PICONEWTON FORCE REALIZATIION SYSTEM

J.-H. Choi, H. H. Choi
  • Abstract:
    We present a preparation procedure for loading flux quantum weights on piconewton force standard system, being developed at KRISS, including the micropositioning of a cantilever device with a superconducting micro-ring. To locate the device at zero-magnetic-field position inside a high-field-gradient superconducting magnet, the behaviour of the mechanical resonance frequency of the device in magnetic fields were studied in a dynamic mode and used to probe the local magnetic field value at the device position. A vertical offset of a few hundred micrometers caused a shift of ~ 0.7 Hz to the resonance frequency of 1144.4 Hz, and was nulled by either additional uniform magnetic field or a z-axis nanopositioner. Nulling was also confirmed by a non-singular temperature dependence of the resonance frequency.
  • Keywords:
    micro-position, flux quantum, piconewton, force standard
  • DOI:
    _unreg_wc-2012.TC3-O

Event details:

  • IMEKO TC:
  • Event name:
    XX IMEKO World Congress
  • Title:

    Metrology for Green Growth

  • Place:
    Busan, REPUBLIC of KOREA
  • Time:
    09 September 2012 - 12 September 2012