Skip to main content

PREDICTION OF THE OUTPUT VOLTAGE OF DC VOLTAGE STANDARDS

Damir Ilic, Alan Šala, Ivan Lenicek
  • Abstract:
    In this paper we report some methods which upgrade the measurement uncertainty evaluation in analysis of the prediction of the output voltage of DC Reference Standards (Fluke 732A at 1 V, 1.018 V and 10 V levels). The monitoring of long-term stability of Zener-based voltage reference standards is one of the main tasks of Primary Electromagnetic Laboratory (PEL), which is a part of the Faculty of Electrical Engineering and Computing of the University of Zagreb, and a holder of national standards of voltage in Croatia. The long-term stability of its DC Reference Standard has been analysed according to the calibration data obtained in different laboratories over a period of more than twenty years. The weighted least-squares fitting was used and the regression coefficients were calculated, by which the maintained voltages at all three outputs can be predicted for a moment of interest. This approach significantly reduces the uncertainties of the maintained voltages.
  • Keywords:
    DC reference voltage standard, least-squares analysis, long-term drift
  • DOI:
    _unreg_wc-2009.670

Event details:

  • IMEKO TC:
  • Event name:
    XIX IMEKO World Congress
  • Title:

    Fundamental and Applied Metrology

  • Place:
    Lisbon, PORTUGAL
  • Time:
    06 September 2009 - 11 September 2009