Skip to main content

pH Sensor Calibration Procedure

Carla Artero-Delgado, Marc Nogueras-Cervera, Antoni Mànuel-Làzaro
  • Abstract:
    This paper describes the calibration of pH sensor located at the OBSEA marine observatory. This instrument is based on an industrial pH electrode that is connected to a CTD instrument (Conductivity, Temperature, and Depth). The calibration of the pH sensor has been done using a high precision spectrophotometer pH meter from Institute of Marine Sciences (ICM), and in this way it has been obtained a numerical function for the pH sensor proportional to real pH.
  • Keywords:
    pH sensor, CTD, Calibration, OBSEA
  • DOI:
    _unreg_tc4-2013.168

Event details:

  • IMEKO TC:
    TC4
  • Event name:
    TC4 Symposium 2013
  • Title:
    19th IMEKO TC4 Symposium Measurements of Electrical Quantities (together with 17th TC4 IWADC Workshop on ADC and DAC Modelling and Testing)
  • Place:
    Barcelona, SPAIN
  • Time:
    18 July 2013 - 19 July 2013