Search Proceedings
Who we are
PERFORMANCE IMPROVEMENT OF LIQUID FLOW CALIBRATORS BY APPLYING SPECIAL MEASUREMENT AND CONTROL STRATEGIES
Rainer Engel, Hans-Joachim Baade, Andreas Rubel
-
Abstract:PTB?s new "Hydrodynamic Test Field", which represents a high-accuracy water flow calibration facility, will serve as the national primary standard for flow measurands: volumetric and mass flow rate, respectively, and total flow measurement, i.e. the quantity of fluid (volume or mass) passing a flowmeter. Owing to this application aspect, the main design goal was to realize a total expanded measurement uncertainty as low as 0,02 % for total volumetric flow-rate measurement. To meet this decisive requirement, low-uncertainty components were combined with state-of-the-art measurement and control strategies.
-
Keywords:
-
Download:
-
DOI:_unreg_tc9-2003.068
Event details:
-
IMEKO TC:TC9
-
Event name:FLOMEKO 2003
-
Title:11th Conference on Flow Measurement
-
Place:Groningen, NETHERLANDS
-
Time:12 May 2003 - 14 May 2003