Skip to main content

LOW TEMPERATURE CALIBRATION FACILITIES AT KRISS

Inseok Yang, Yong-Gyoo Kim, Chang Ho Song, Kee Hoon Kang, Kee Sool Gam
  • Abstract:
    We review the low temperature calibration facilities at the Korea Research Institute of Standards and Science. For the calibration of capsule-type standard resistance thermometers, open and sealed cells of cryogenic fixed points and their realization system were manufactured and used. A sealed cell of the triple point of Argon for long-stem standard platinum resistance thermometers was also manufactured. For calibration by comparison, a controllable comparison cryostat and a nitrogen boiling point temperature block were made. We also review our traceability scheme and supporting international key comparisons for the dissemination of low temperature standards.
  • Keywords:
    low temperature, cryogenic fixed point, calibration
  • DOI:
    _unreg_wc-2009.547

Event details:

  • IMEKO TC:
  • Event name:
    XIX IMEKO World Congress
  • Title:

    Fundamental and Applied Metrology

  • Place:
    Lisbon, PORTUGAL
  • Time:
    06 September 2009 - 11 September 2009