Search Proceedings
Who we are
LOW TEMPERATURE CALIBRATION FACILITIES AT KRISS
Inseok Yang, Yong-Gyoo Kim, Chang Ho Song, Kee Hoon Kang, Kee Sool Gam
-
Abstract:We review the low temperature calibration facilities at the Korea Research Institute of Standards and Science. For the calibration of capsule-type standard resistance thermometers, open and sealed cells of cryogenic fixed points and their realization system were manufactured and used. A sealed cell of the triple point of Argon for long-stem standard platinum resistance thermometers was also manufactured. For calibration by comparison, a controllable comparison cryostat and a nitrogen boiling point temperature block were made. We also review our traceability scheme and supporting international key comparisons for the dissemination of low temperature standards.
-
Keywords:low temperature, cryogenic fixed point, calibration
-
Download:
-
DOI:_unreg_wc-2009.547
Event details:
-
IMEKO TC:
-
Event name:XIX IMEKO World Congress
-
Title:
Fundamental and Applied Metrology
-
Place:Lisbon, PORTUGAL
-
Time:06 September 2009 - 11 September 2009