Search Proceedings
Who we are
INFLUENCE OF EXCITATION VOLTAGE OVER METROLOGICAL PERFORMNACE OF FORCE TRANSDUCERS
Harish Kumar, Anil Kumar
-
Abstract:The present study discusses the effect of excitation voltage of the digital indicator over the metrological performance of the force transducers. The force transducers of 1 kN, 2 kN and 5 kN nominal capacity and having relative repeatability error better than ± 0.002 % have been calibrated on the basis of calibration procedure based on ISO 376-2004. The relative deviations due to repeatability and reproducibility error have been computed for the force transducers and results have been reported.
-
Keywords:force transducer, excitation voltage, ISO 376-2004, metrological performance
-
Download:
-
DOI:_unreg_wc-2012.TC3-P11
Event details:
-
IMEKO TC:
-
Event name:XX IMEKO World Congress
-
Title:
Metrology for Green Growth
-
Place:Busan, REPUBLIC of KOREA
-
Time:09 September 2012 - 12 September 2012