  • Abstract:
    The present study discusses the effect of excitation voltage of the digital indicator over the metrological performance of the force transducers. The force transducers of 1 kN, 2 kN and 5 kN nominal capacity and having relative repeatability error better than ± 0.002 % have been calibrated on the basis of calibration procedure based on ISO 376-2004. The relative deviations due to repeatability and reproducibility error have been computed for the force transducers and results have been reported.
  • Keywords:
    force transducer, excitation voltage, ISO 376-2004, metrological performance