IMPROVEMENT OF VERY LOW-FREQUENCY PRIMARY VIBRATION CALIBRATION SYSTEM AT NMIJ/AIST

W. Kokuyama, T. Ishigami, H. Nozato, A. Ota
  • Abstract:
    The very low-frequency primary vibration calibration system at NMIJ/AIST was improved with a new signal processing framework. It contains commercial hard disk drive so that photoelectric signals from homodyne interferometer can be recorded. Sine-approximation method with a digital band-pass filter is applied to boost signal-to-noise ratio. By calibrating a reference accelerometer, compatibility with the existing primary vibration calibration system was confirmed.
  • Keywords:
    very low frequency, primary vibration calibration, band-pass filter, sine approximation method