Skip to main content

IMPACT OF AC ELECTRIC FIELD NON-UNIFORMITY ON IMPEDANCE OF CONDUCTIVITY CELL

A. A. Mikhal, Z. L. Warsza
  • Abstract:
    AC impedance model of a liquid conductivity has been developed to assess the influence of magnetic vector potential. It has been proved that the field in a liquid conductor is quasi-static. Approximation error, even taken for very concentrated solutions (e.g. sea water), does not exceed 0.01 ppm. The calculation of resistance of the four-electrode cell is given with the account of Ampere's circuit law. Two types of errors are defined: DC error due to finite thickness of electrodes, and AC error due to presence of displacement currents when using an impedance model.
  • Keywords:
    measurement, electrolytic conductivity, electrical field, error
  • DOI:
    _unreg_wc-2015.460

Event details:

  • IMEKO TC:
  • Event name:
    XXI IMEKO World Congress
  • Title:

    Measurement in Research and Industry

  • Place:
    Prague, CZECH REPUBLIC
  • Time:
    30 August 2015 - 04 September 2015