Jaroslav Bohácek
-
Abstract:With an intention to reveal possible inaccuracies of mathematical models used in calculations of frequency characteristics of resistors of octofilar design, comparisons of a 12 906 Ω and a 1 Ω octofilar resistor against monofilar and quadrifilar calculable resistors of the same nominal values have been performed. A possibility to carry out these precision comparisons by means of a commercially available LCR meter has been investigated.
-
Keywords:calculable resistor, octofilar design, LCR meter
-