Search Proceedings
Who we are
EURAMET cg-18 - Calibration of Balances and Scales - Update on International Distribution and Acceptance
Klaus Fritsch
-
Abstract:This paper presents the state-of-the-art calibration strategy for non-automatic weighing instruments (NAWI), based on the recently revised guideline EURAMET cg-18. Besides of the uncertainty at calibration, the document provides advice for the so-called uncertainty of a weighing result that describes the device performance during day-to-day use. The concept of minimum weight was added during a recent revision. The minimum weight is the smallest sample quantity that can be weighed on the device, still adhering to a predefined relative weighing tolerance requirement.
-
Keywords:calibration, non-automatic weighing instruments, minimum weight, harmonization, quality
-
Download:
-
DOI:_unreg_tc3-2017.008