Skip to main content

ELECTRONIC SKIN REPLACES VOLUNTEER EXPERIMENTS

Norbert Nessler, Werner Reischer, Marcel Salchner
  • Abstract:
    Neutral electrodes (grounding plates used in electrosurgical operations) were tested with thermo cameras in volunteers experiments. These experiments are difficult to perform and expensive. A new test device swaroTEST with a measuring board "electronic skin" simulates the relevant electrical features of human skin to replace volunteer experiments. This device consists of a three dimensional resistor network representing the electric features of the skin and muscle tissue, and a temperature sensing array (one transistor for each cm²) to measure the resultant temperature increase after a standardized current load (700 mA hf current during 60 s, proposed in the relevant AAMI HF-18 standard). The comparison of test results with thermo camera images of volunteer experiments showed a sufficient coincidence, which proves the applicability of this test device to replace volunteer experiments.
  • Keywords:
    electronic skin, electrosurgery, neutral electrode
  • DOI:
    _unreg_wc-2003.TC10-006

Event details:

  • IMEKO TC:
  • Event name:
    XVII IMEKO World Congress
  • Title:

    Metrology in the 3rd Millennium

  • Place:
    Dubrovnik, CROATIA
  • Time:
    22 June 2003 - 28 June 2003