Hideaki Iizumi, Hiroaki Kajikawa, Tokihiko Kobata, Kenichi Fujii
  • Abstract:
    The National Metrology Institute of Japan, AIST (NMIJ/AIST) has developed the high gas pressure standard for pressures up to 100 MPa for more reliable pressure measurements. A high gas pressure calibration system using a liquid-lubricated pressure balance has been developed. The effective area of the piston-cylinder assembly is determined from the comparison against hydraulic pressure balances. Calibration results of a pressure transducer are compared with those using the hydraulic pressure standard.
  • Keywords:
    pressure standard, high gas pressure, liquid-lubricated pressure balance, calibration system
  • DOI:
    _unreg_wc-2015.334