Search Proceedings
Who we are
DETERMINATION OF FORM MEASURING MACHINE DISPLACEMENT SENSOR CHARACTERISTICS WITH A USE OF FLICK STANDARD
Marta Wiśniewska, Sabina Żebrowska-Łucyk
-
Abstract:A new approach towards calibration of FMM displacement sensor with a use of a flick standard is presented. The main drawback of the prevailing procedure is the fact that instrument’s probe could have been calibrated only for a single value, whereas following a novel method removes this limitation. Therefore, FMM probe linearity deviations may be determined within a full range of sensor. In the paper, the core idea and sample results presenting performance of the proposed method are outlined.
-
Keywords:form measuring machine, calibration, sensor characteristics, flick standard
-
Download:
-
DOI:_unreg_wc-2015.320
Event details:
-
IMEKO TC:
-
Event name:XXI IMEKO World Congress
-
Title:
Measurement in Research and Industry
-
Place:Prague, CZECH REPUBLIC
-
Time:30 August 2015 - 04 September 2015