Skip to main content

Design of Torque Standard Machine for Calibration of Reference Torque Wrench at NIM

Meng Feng, Zhang Zhimin, Li Tao, Lin Jing
  • Abstract:
    The paper describes the design of a new torque standard machine at NIM. This machine is used to calibrate the reference torque wrench and torque transducer. The measurement uncertainty is expected to reach 0.03 % (k = 2). The range is designed from 10 N · m to 5000 N · m.
  • Keywords:
    torque standard machine, reference torque wrench, calibration method
  • DOI:
    _unreg_tc3-2017.062

Event details:

  • IMEKO TC:
    TC3
  • Event name:
    TC3 Conference 2017 - Measurement facing new challenges!
  • Title:

    23rd Conference on the Measurement of Force, Mass and Torque (together with 4th TC22 Conference on Vibration Measurement and 13th TC5 Conference on the Measurement of Hardness)

  • Place:
    Helsinki, FINLAND
  • Time:
    30 May 2017 - 1 June 2017