DESIGN OF HELIUM PERMEATION REFERENCE LEAKS WITH GENERATED FLOWS BELOW 10-14 mol/s

Bojan Erjavec, Janez Šetina
  • Abstract:
    The calibration of helium permeation reference leaks below 10-14 mol/s is presented. The reference leak is based on the known construction with a He reservoir and a permeable glass tube leak element. The leak rate is determined in situ using the rate of pressure rise method. The reference leak is equipped with a small calibration volume containing a miniature Ti-getter pump to reduce the residual pressure rise due to outgassing and a spinning rotor gauge (SRG) to measure the rate of He pressure rise. In vacuum metrology, the SRG has been recognized as a transfer standard due to its inertness, high resolution and excellent long-term stability. The He leak rate ranging from 4,0×10-17 mol/s to 1,0×10-14 mol/s, measured at the reference leak temperature of 23°C, is linearly dependent on the He fill pressure ranging from 4,6 Pa up to 1120 Pa. Multiple He flow measurements in the 10-15 mol/s range show that a measurement uncertainty less than 2% can be achieved.
  • Keywords:
    gas permeation, gas flow, spinning rotor gauge, helium reference leak
  • DOI:
    _unreg_wc-2003.TC1-021

Event details:

  • IMEKO TC:
  • Event name:
    XVII IMEKO World Congress
  • Title:

    Metrology in the 3rd Millennium

  • Place:
    Dubrovnik, CROATIA
  • Time:
    22 June 2003 - 28 June 2003