Skip to main content

DESIGN, DEVELOPMENT AND ESTABLISHMENT OF 1 kN FORCE CALIBRATION MACHINE IN FORCE LABORATORY OF TSE

Bulent Aydemir, Sinan Fank, Cemal Vatan, Dervis Anil, Alican Kuyrukluyildiz
  • Abstract:
    A new dead weight type force calibration machine with a 1 kN loading capacity both in tension and compression is installed both in force laboratory of UME and TSE. This paper describes the preliminary results of the performance of the newly developed 1 kN dead weight force machine at the Force Laboratory. The machine is designed and manufactured by UME. The performance test results show that relative reproducibility error with rotation (b) ± 5,5 × 10-3 and relative repeatability error without rotation (b’) ± 3,5 × 10-3 for a range of 0,2 kN to 1 kN of a new 1 kN machine of TSE.
  • Keywords:
    force calibration machine, dead weight, performance test
  • DOI:
    _unreg_wc-2015.029

Event details:

  • IMEKO TC:
  • Event name:
    XXI IMEKO World Congress
  • Title:

    Measurement in Research and Industry

  • Place:
    Prague, CZECH REPUBLIC
  • Time:
    30 August 2015 - 04 September 2015