Skip to main content

DEFORMATION COMPENSATION OF A STABILIZED RESONATOR UNDER HIGH PRESSURE CHANGE BASED ON A MODIFIED POUND-DREVER-HALL METHOD

Duong Quang Anh, Tomokage Syunsaku, Masato Aketagawa
  • Abstract:
    Recently, the resolution of length measurements based on optical interferometry reached sub-nanometer order. However, the accuracy of the optical interferometry is limited by the knowledge of absolute air refractive index. In this work, we have proposed a method to measure the absolute air refractive index by measuring the free-spectral-range and the resonance frequency of a Fabry-Perot cavity, whose inside is filled by vacuum and by air. During the air refractive index measurement, the geometrical length of the Fabry-Perot cavity must be constant in vacuum and in air. In this paper, a compensation of the Fabry-Perot cavity length deformation owing to a large pressure difference between air and vacuum is proposed.
  • Keywords:
    air refractive index, Fabry-Perot cavity, Free spectral range, Resonance frequency, Pound-Drever-Hall method
  • DOI:
    _unreg_tc14-2014.08

Event details:

  • IMEKO TC:
    TC14
  • Event name:
    TC14 LMPMI Symposium 2014
  • Title:

    11th Symposium on Laser Metrology for Precision Measurement and Inspection in Industry

  • Place:
    Tsukuba, JAPAN
  • Time: