Skip to main content

CALIBRATION OF MULTI-DECADE INDUCTIVE VOLTAGE DIVIDERS

J. Bohacek, M. Andrle
  • Abstract:
    A method for calibrating multi-decade inductive voltage dividers has been developed which makes it possible to determine both the inphase and the quadrature error for ratios of i /11, i = 1, 2, ..., 10 in the frequency range from 400 Hz to 5 kHz. Calibration is made by comparing the divider under test against a reference two-stage divider having a ratio winding with 11 identical sections. In calibration of the reference divider itself, an auxiliary 11:1 transformer is used.
  • Keywords:
    inductive voltage divider, calibration, ratio error
  • DOI:
    _unreg_wc-2000.216

Event details:

  • IMEKO TC:
  • Event name:
    XVI IMEKO World Congress
  • Title:

    Measurement - Supports Science - Improves Technology - Protects Environment ... and Provides Employment - Now and in the Future

  • Place:
    Vienna, AUSTRIA
  • Time:
    25 September 2000 - 28 September 2000