Search Proceedings
Who we are
Calibration of low-current meters at nA level with the capacitance-charging method
Luca Callegaro, Vincenzo D'Elia
-
Abstract:The most accurate generation of dc currents in the lower ranges is performed with the capacitance-charging technique. Implementations of the method are typically limited to an upper current in the 100 pA range. Here we present a range extender which can be employed as plug-in in existing setups. The extender permits to scale the current of a ×10 factor and reach the nA range. Results of calibration of low-current meters with and without the use of the extender are reported.
-
Keywords:low current, calibration, picoammeter, dc current, capacitance
-
Download:
-
DOI:_unreg_tc4-2013.016
Event details:
-
IMEKO TC:TC4
-
Event name:TC4 Symposium 2013
-
Title:19th IMEKO TC4 Symposium Measurements of Electrical Quantities (together with 17th TC4 IWADC Workshop on ADC and DAC Modelling and Testing)
-
Place:Barcelona, SPAIN
-
Time:18 July 2013 - 19 July 2013