Search Proceedings
Who we are
CALIBRATION OF FIBER OPTIC POWER METER AT 1625 nm
Seung Kwan Kim, Kee Suk Hong, Ki-Lyong Jeong, Yong Seok Kim
-
Abstract:Fiber optic power responsivity scale at 1625 nm is newly established to support optical fiber communication industry. Using the absolute scale at reference level of -10 dBm and the linearity measurements, a fiber optic power meter is calibrated at 1625 nm from -60 dBm to 0 dBm. The uncertainty is less than 0.7 % (k = 2) at -10 dBm and about 1 % in the full range.
-
Keywords:fiber optic power, responsivity, calibration, fiber optic power meter
-
Download:
-
DOI:_unreg_wc-2015.018
Event details:
-
IMEKO TC:
-
Event name:XXI IMEKO World Congress
-
Title:
Measurement in Research and Industry
-
Place:Prague, CZECH REPUBLIC
-
Time:30 August 2015 - 04 September 2015