Skip to main content

AN ECONOMIC WAY TO HAVE SMALL TEMPERATURE VARIATION CONTROLLED IN MASS MEASUREMENT

Feng-yu Yang, Sheau-shi Pan, Sheng-Jui Chen
  • Abstract:
    In APMF2011, we introduced a simple way by building a two-layer insulation chamber with two mass comparators inside to have small temperature variation in our laboratory. In this paper, we will introduce a economic method by adding a simple active control unit on the aluminum chamber to control the temperature inside the chamber at (20 ± 0.1) °C nearly, while the room temperature of the air-conditioning laboratory is active controlled at (21.5 ± 1.5) °C. Using this method, the transient phenomenon of temperature was observed.
  • Keywords:
    temperature variation / fluctuation, passive / active control, two-layer insulation chamber
  • DOI:
    _unreg_wc-2012.TC3-P22

Event details:

  • IMEKO TC:
  • Event name:
    XX IMEKO World Congress
  • Title:

    Metrology for Green Growth

  • Place:
    Busan, REPUBLIC of KOREA
  • Time:
    09 September 2012 - 12 September 2012