Search Proceedings
Who we are
AN ECONOMIC WAY TO HAVE SMALL TEMPERATURE VARIATION CONTROLLED IN MASS MEASUREMENT
Feng-yu Yang, Sheau-shi Pan, Sheng-Jui Chen
-
Abstract:In APMF2011, we introduced a simple way by building a two-layer insulation chamber with two mass comparators inside to have small temperature variation in our laboratory. In this paper, we will introduce a economic method by adding a simple active control unit on the aluminum chamber to control the temperature inside the chamber at (20 ± 0.1) °C nearly, while the room temperature of the air-conditioning laboratory is active controlled at (21.5 ± 1.5) °C. Using this method, the transient phenomenon of temperature was observed.
-
Keywords:temperature variation / fluctuation, passive / active control, two-layer insulation chamber
-
Download:
-
DOI:_unreg_wc-2012.TC3-P22
Event details:
-
IMEKO TC:
-
Event name:XX IMEKO World Congress
-
Title:
Metrology for Green Growth
-
Place:Busan, REPUBLIC of KOREA
-
Time:09 September 2012 - 12 September 2012