Skip to main content

Accurate Measurement of the RMS Value by Means of an Analog Multiplier-based RMS-to-DC Converter

Daniel Belega, Robert Pazsitka, Dan Stoiciu
  • Abstract:
    This paper investigates the influence of the accuracy of the op-amp used by the square root circuit on the accuracy of the root mean square (rms) measurement achieved by means of an analog multiplier-based rms-to-dc converter implemented using the direct method. The rms measurement accuracy achieved using the internal op-amp of the multiplier is compared through experimental results with that achieved using more accurate op-amps, namely TL082 and OPA277. The analyzed rms-to-dc converter is implemented using the Texas Instruments Analog System Lab Kit (ASLK) PRO board equipped with MPY64 analog multipliers. The experimental data are obtained by means of a dedicated experimental setup.
  • Keywords:
    analog multiplier, error analysis, op-amp, rms-to-dc converter
  • DOI:
    _unreg_tc4-2017.068

Event details:

  • IMEKO TC:
    TC4
  • Event name:
    TC4 Symposium 2017
  • Title:

    22nd IMEKO TC4 Symposium and 20th International Workshop on ADC Modelling and Testing
    "Supporting world development through electric&electronic measurements"

  • Place:
    Iasi, ROMANIA
  • Time: