Search Proceedings
Who we are
A NEW CALIBRATION DEVICE FOR SILK SCREEN TENSION METER
Hu Xiaoyin, Yu Yueling, Yu Peiying, Kang Zhangy
-
Abstract:In this paper, a calibration device for silk screen tension meter is designed. The device has a measurement range from 5 to 100 N/cm. It can achieve simultaneous loading in x and y directions by servo motors. For loading accuracy, two high precisionforce sensors are utilized in the closed-loop control system for each direction. To verify the device, it is utilized to calibrate a commercial available silk screen tension meter. The results from the device are compared with that from the Swiss Federal Office of Metrology. It demonstrates that the max error is 0.5 %.
-
Keywords:silk screen tension meter, calibration device, force, DIN 16611-1990
-
Download:
-
DOI:_unreg_tc3-2014.037
Event details:
-
IMEKO TC:TC3
-
Event name:TC3 Conference 2014
-
Title:
22nd Conference on the Measurement of Force, Mass and Torque (together with 3rd TC22 Conference on Vibration Measurement and 12th TC5 Conference on the Measurement of Hardness)
-
Place:Cape Town, SOUTH AFRICA
-
Time:03 February 2014 - 06 February 2014