Skip to main content

A METHOD OF TRACEABILITY FOR A FPG8601 FORCE BALANCED PISTON GAUGE TO DEFINE PRESSURES IN THE RANGE FROM 1 PA TO 15 KPA IN GAUGE AND ABSOLUTE MEASUREMENT MODES

Rob Haines, Michael Bair
  • Abstract:
    The FPG8601 has become an important reference for many primary measurement laboratories in a pressure range from approximately 1 Pa to 15 kPa in both gauge and absolute measurement modes. The FPG8601 measurement range spans the gap between the traditional measurement regimes of "pressure" and "'vacuum", a range where other primary pressure standards either do not exist or do not have uncertainties sufficient to support the FPG8601 traceability. The limitations of available primary standards create a challenge for FPG8601 users needing to define traceability in this range. This paper suggests a method of maintaining traceability using force, dimensional and primary pressure measurements in a manner different from that normally used for traditional "floating" piston gauges.
  • Keywords:
    pressure, vacuum, traceability
  • DOI:
    _unreg_wc-2009.298

Event details:

  • IMEKO TC:
  • Event name:
    XIX IMEKO World Congress
  • Title:

    Fundamental and Applied Metrology

  • Place:
    Lisbon, PORTUGAL
  • Time:
    06 September 2009 - 11 September 2009