A CURRENT SOURCE FOR CALIBRATION OF LOW-CURRENT METERS

Luca Callegaro, Vincenzo D’Elia
  • Abstract:
    The need for accurate calibration of current meters in the picoampere range is continuously increasing. A traceable current source for calibration of meters in the 100 fA – 100 pA current range is here presented. The source is based on a low-frequency generator ramp generator, which charges and discharges a gas-dielectric capacitor. Current traceability is given by the measurement of voltage parameters with a sampling voltmeter, and by calibration of the capacitor at audio frequency. The source has been employed in the EUROMET.EM-S24 "Comparison of small current sources" supplementary comparison.
  • Keywords:
    current measurement, calibration, signal generators
  • DOI:
    _unreg_wc-2006.TC4-003

Event details:

  • IMEKO TC:
  • Event name:
    XVIII IMEKO World Congress
  • Title:

    Metrology for a Sustainable Development

  • Place:
    Rio de Janeiro, BRAZIL
  • Time:
    17 September 2006 - 22 September 2006