Skip to main content

1 MHZ MODULE TRANSFORMER BRIDGE

M. D. Klionskiy
  • Abstract:
    A transformer bridge to study low-loss dielectrics at the frequency 1 MHz is developed. The main components of the bridge: a ratio transformer and variable measuring standards of capacitance and conductance are made in the form of removable modules. The bridge measurement range by loss angle tangent D is 10-6 - 10-3 with the capacitance 1-1000 pF. Measurement uncertainty D is ± (2-3) &itimes; 10-6, k = 2.
  • Keywords:
    transformer bridge, loss angle
  • DOI:
    _unreg_wc-2000.222

Event details:

  • IMEKO TC:
  • Event name:
    XVI IMEKO World Congress
  • Title:

    Measurement - Supports Science - Improves Technology - Protects Environment ... and Provides Employment - Now and in the Future

  • Place:
    Vienna, AUSTRIA
  • Time:
    25 September 2000 - 28 September 2000