A PRECISION CALIBRATION SET-UP FOR AC MAGNETIC FLUX DENSITY MEASUREMENT IN THE RANGE OF 1 Hz TO 20 kHz
Po Gyu Park, Young Gyun Kim, Wan-Seop Kim, V. N. Kalabin, Vladlen Ya. Shifirn
-
Abstract:A precision generating and measurement system of AC magnetic flux density was developed by using a linear low-frequency AC magnetic field converter, a quartz Helmholtz coil, a voltage comparator, a multimeter, a standard resistor, and a computer with a program for automatic measurement. The expanded uncertainty (k = 2) of the calibration system is less than 0.2 % over the frequency range of 1 Hz to 20 kHz.
-
Keywords:AC magnetic flux density, Calibration
-
Download:
-
DOI:_unreg_wc-2009.197
Event details:
-
IMEKO TC:
-
Event name:XIX IMEKO World Congress
-
Title:
Fundamental and Applied Metrology
-
Place:Lisbon, PORTUGAL
-
Time:6 September 2009 - 11 September 2009